Background
Lead in drinking water is a major global problem. The
sources of the lead are from solder, which was used to join copper pipes,
and from lead pipes used in older homes and buildings. Over time, lead
can leach out and contaminate water. Acute lead poisoning in children
can cause anorexia, vomiting, malaise and convulsions. Permanent brain
damage can result. Chronic lead poisoning can cause weight loss, weakness
and anemia.
The following laboratory procedure will enable you to measure lead at
low concentrations in the water you drink.
Anodic Stripping Voltammetry (ASV)
In recent years there has been a revival of interest
in anodic stripping voltammetry (ASV), which is one of the most sensitive
analytical methods for heavy metals like lead. It can detect lead at
levels as low as few parts per billion (ppb). An advantage of ASV is
its low cost compared to many other methods. ASV involves the preconcentration of lead into a Hg
or Hg film electrode by electro-depositing it at a negative potential.
After deposition, the potential is scanned anodically toward positive
potentials to strip (oxidize) the lead from the electrode back into
the solution. This stripping results in a current peak at the potential
for oxidizing the lead. If more than one metal is deposited, multiple
peaks are found on the anodic potential scan. In the experiment to follow,
a "bare" glassy carbon electrode is used to eliminate any
exposure to Hg, which is toxic if ingested.
Experimental
Equipment:
- 66-CS1090 or 66-CS1200
computer-controlled potentiostat or 66-OMNI101 microprocessor controlled
potentiostat with x-y or strip-chart recorder
- Electrochemical cell
fitted with a 1.5 mm diameter glassy carbon electrode (working electrode),
- A Pt auxiliary (counter)
electrode and an Ag/AgCl reference electrode
- A small magnetic stirring
bar inside the cell
- A magnetic stirrer
Chemical solutions:
- Prepare three separate
solutions containing 100 ppb, 500 ppb and 1,000 ppb solutions of
Pb(NO3)2 in 0.1 M KNO3 containing 50 mM HNO3 (use doubly
distilled water for dilutions)
- Dilute a sample of
tap water diluted with an equal volume of 0.1 M KNO3 containing
50 mM HNO3
Procedure:
With the 66-Omni101-
- Connect an x-y recorder or strip-chart
recorder to the 66-Omni101 (or computer with data acquisition
software, ACQUIRE 101).
- Set the cell on a magnetic stirrer and
put the auxiliary and reference electrodes in place
- Lightly polish the glassy carbon electrode
with an alumina slurry on a microcloth polishing pad using a figure
8 motion. Rinse the electrode quickly with doubly distilled water,
remove any excess water by touching an edge with Kimwipe and fit
the electrode into the cell
- Add the 100 ppb Pb sample in the cell
and replace the cover with the electrodes
- Set the applied potential, Ei,
at -900 mV and final potential, Ef, to + 100 mV.
- Turn the magnetic stirrer ON, set the
time for 2 minutes and then switch the potentiostat ON so that the
-900 mV is applied to the glassy carbon electrode vs the Ag/AgCl
reference
- If the current overload light turns
on, decrease the sensitivity.
- After 1 minute and 30 seconds, turn
the stirrer off so that the solution quiets
- When the timer goes off marking 2 minutes,
push the scan button ON so that the potential is scanned from Ei
of -900 mV to Efof +100 mV. A scan rate of 500 mV/sec
is recommended to obtain a good peak height (a lower scan rate gives
smaller peak heights and also allows possible interference from
oxygen)
- Repeat the experiment with all three
standard solutions and then run the tap water sample
- Plot the peak heights versus concentration
for the standards and then determine the amount in the tap water
from the calibration plot (where the peak height of the sample lies
on plot for the standards)
- Do a linear regression analysis to confirm
the ppb lead in the tap water sample (run at least duplicates on
the standards and the sample if time permits)
If you use the 66-CS1090 or 66-CS1200 computer-controlled
potentiostat the procedure for the analysis of lead is the same as that
described above. You can use scan rates as high as 1 or 2 V/s with these
instruments. The computer will do all the timing and potential setting/scanning
but please remember to turn the stirrer off at the preset time of 2
minutes. A scan rate of 1.0 or 2.0 V/s is recommended. Additional Experiments
- Instead of electrodepositing
lead directly on the glassy carbon electrode, you can dissolve
Hg(NO3)2 in the solutions so that Hg is co-deposited along with the Pb to make
an Hg/Pb amalgam. If you do this, you may see peaks for copper and
cadmium if present in the solution.
- You can use standard
addition methods to calibrate for metals of interest. The solution
pH may need adjustment to optimize results with some metals.
- ASV is an excellent
method for heavy metal analysis and comparison of results to traditional
wet chemical methods may be of interest to students.
Reference for ASV Wang, Joseph, Anodic Stripping Voltammetry: An
Instrumental Analysis Experiment, J. Chem. Ed. 60, December 1983,
pages 1074-1075. Additional lab experiments are being developed for
the undergraduate teaching laboratory to introduce students to electroanalytical
chemistry. |