66-CS1200 Computer Control Potentiostat

Ideal for:

Conducting a wide range of electroanalytical experiments
Controlling/monitoring other devices

Cypress System's 66-CS1200 Computer Control Potentiostat

Introduction:

   The 66-CS1200 is a high sensitivity (<100 pA) and high scan rate (up to 250 V/s) potentiostat that uses a National Instruments AD/DA interface for reliability, versatility and Windows compatibility without obsolescence. To design and control your own experiments, the 66-CS1200 has two analog input channels (BNC connectors) for monitoring signals from other sources such as pH or ISE electrodes, temperature, etc). There is a TTL trigger output channel for timed start/stop of devices such a stirrer, flow pump or recorder. Two additional BNC connectors on the front panel allow analog monitoring of the cell potential and current with any XY or XYT recorder or oscilloscope. A DB9F connector on the back panel provides two digital I/O control lines, one trigger input line for starting the potentiostat "ON" externally, and an opto-electronic output trigger signal for linking to higher power devices. Cypress’ custom software, based on NI platform, is pre-loaded and supplied with a WIN 2000 or XP computer. Inquire about customized repetitive pulse waveform methods.

Features:

High sensitivity to 50 pA
High speed, scan rates up to 250 V/s
Computer controlled
Wide current range up to 40 mA @ 16V
Programmable waveforms to tailor your experiments
TTL outputs to trigger other devices (e.g., pumps and valves)
Inputs to monitor signals from other devices
Import data files (ASCII format) to other programs like Excel ®

Techniques:

Cyclic voltammetry
Linear sweep voltammetry
Differential pulse voltammetry
Square wave voltammetry
Chronoamperometry
Chronocoulometry
Anodic stripping voltammetry
Controlled potential electrolysis

Applications and Experiments:

Cyclic Voltammetry
Determine kinetic rate of an ec mechanism
Study of redox properties – Bioinorganic systems
Environmental Analysis
Pulse voltammetric methods
Anodic Stripping Voltammetry
Fast cyclic voltammetry

Options:

Computer automated electrochemical flow analysis system
Pre-loaded Excel program
Laptop computer with Nat’l PCMIA Interface

Software

 Software data system
* Please call for software options.

Specifications:

Compliance voltage 16 V
Compliance current 40 mA
Applied potential 2 V or 4 V
Data acquisition rate 100 kHz throughput
D/A 12 Bit, 1 mV resolution minimum
A/D 12 Bit, 0.024% of full scale range
Rise time 10 µs, depends on sensitivity setting
Current ranges 500 pA to 50 mA
Minimum current 50 pA
Maximum current 40 mA
Physical dimensions 8" x 11" x 2", 2 lbs